SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
How to use Frequency Counter
— Press Marker.
— Press Normal, Delta, or Peak, or select an existing marker to use as a
— Locate the marker somewhere on a CW signal. It does NOT have to be on
— Then
— Then
— With Freq Count ON, you can scroll through the existing markers that are

Time Zero Fixed Marker

This feature was created to allow recall of vintage instrument states (older than
Rev. 7.0) that included Zero span sweep with a trigger delay and at least one
marker. Before Rev. 7.0, these instrument states were saved and recalled with
the equivalent of the ON state of this setting. The default setting is OFF. Learn
more about Triggering with negative delay in
— Press Mkr->/Tools
— Then
— Then
— Then
— Then choose from the following:

Audio Beep at Marker Power

When enabled, the Audio Beep feature emits a repetitive beep sound which
varies in tone pitch and repetition rate to indicate the relative power level of the
active marker. The highest tone pitch and fastest beep rate occurs when the
marker Y-axis position is at the top of the display. Conversely, the lowest pitch
and repetition rate occurs at the bottom of the display. Therefore, it is
important to scale the signals that you intend to measure between the top and
bottom of the screen. Learn more about Scale in
page
156
frequency counter.
the peak.
More
Freq Count ON
located on a discrete CW signal, and read the frequency of the signal peak.
However, each frequency count update requires another sweep. If the
FieldFox is in HOLD, press
More
Mkr Properties
Time Zero Fixed
ON
Time Zero is fixed at the left-edge of the screen. A marker in
negative delay resides at this point. This is the vintage behavior.
OFF
(default) Time Zero is located at the T-zero indicators. A marker
in negative delay appears to the left of T-zero.
115.
Single
to update the frequency count results.
"Triggering" on page
"Scale and Units" on
Keysight N9938-90003 User's Guide
136.