Saving and Recalling Files

Saving and Recalling Files

FieldFox State files are intended to be recalled on the same instrument
model on which they were saved. Sometimes you may able to recall them
within model families. An error will result when trying to recall a state file
that was saved on a different model family.
The FieldFox model families are:
— N991x (VNA/SA) model family
— N992x (VNA) model family
— N993x (SA) model family
— N995x (VNA/SA) model family
— N996x (SA) model family

Save Files

How to Save a File
— Press Save/Recall 9
— Then
— At the top of the current 'labeler' screen (see following image) check the
— Then use the labeler to specify a filename.
If you are having difficulty saving to a USB flash drive, it may be a timing
issue. The turn-around time for "file open, write data, file close" to a USB
drive is much longer than to FieldFox's internal memory. Alternative
methods are:
1.
2.
266
Save
File Type to save and Device (Internal, USB, Storage Card) to which the file
is saved. Learn how to change these settings in
Device" on page
271.
Save your content to the FieldFox's internal memory; and then move
your files off to a USB flash drive using FTP SCPI programming. Refer to
Using FTP with FieldFox
Save your data to the FieldFox's internal memory, then move the data
to a PC using a binary block transfer, and delete the data from the
FieldFox's internal memory. An example for transferring an image:
http://na.support.keysight.com/fieldfox/help/Programming/webhelp/
Examples/Transfer_Image_to_PC.htm. Please note that the receiving
computer program needs to be able to handle the binary block transfer. See
also
FieldFox Programming
"Set File Type and Select
and to the
FieldFox Programming
Guide.
Keysight N9938-90003 User's Guide
Guide.