Enhanced Response Optimization - Keysight B Series User Manual

Fieldfox analyzers
Table of Contents

Advertisement

Calibration for NA, CAT, and VVM Modes
— Press Done, then press Back.
— Cable Correction = Auto is the default setting. The Effective Velocity Factor
NA Mode - How to make Waveguide settings
These settings are necessary ONLY when your measurement requires electrical
delay or port extensions, or if using Time Domain Transform.
— Press Meas Setup 4
— Then
— Then
— Under Transform Stimulus Settings, set the Start and Stop frequencies to
— Set Stimulus = Bandpass Impulse
— Press
— Press Meas Setup 4 or Back
— Then
— Set Media = Waveguide
— Set Cutoff Frequency. This is the absolute minimum frequency of the
Enter the calculated Effective VF value into the Velocity Factor setting.

Enhanced Response Optimization

When Enhanced Response Cal Type is selected, either for one calibration or for
CalReady, this setting optimizes the calibration based on the type of DUT
being measured. See also: CalReady Properties in
This setting does NOT survive Preset.
— Press Cal 5 then
— Then
102
— Select User Waveguide. Then press Done.
— Then
DTF Cable
Edit
Cable.
— Scroll to set Waveguide Definition. The default setting is VF Corr =
Auto.
— Set the Min, Max, and Cutoff Frequencies.
is calculated automatically based on the frequencies of the waveguide
standard. To override this setting, set Cable Corr = Man.
Transform
Transform Settings
those of the Waveguide.
Done
Calibration Settings
waveguide. This value must be less than the Start Frequency of the
Waveguide.
More
Enh.Response
Specifications, then
Edit/Save/Recall Cables,
Keysight N9938-90003 User's Guide
"CalReady" on page
89.