SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings

Single / Continuous / Restart

This setting determines whether the FieldFox measures continuously or only
once each time the
Single or to conserve battery power or to allow you to save or analyze a
specific trace.
How to set Single or Continuous
— Press Sweep 3.
— Then choose one of the following:

Points

Points is the number of measurements that are displayed along the X-axis. The
higher number of data points, the better the ability to resolve closely spaced
signals and the slower the sweep speed.
How to set number of points
— Press Sweep 3
— Then
— Then choose from:

Trace Display States (SA Mode)

— In SA Mode you can display up to four of the following types of trace states.
Keysight N9938-90003 User's Guide
Single
—Continuous
——ON
Causes the SA to make continuous sweeps. This is the
typical setting for making measurements. Each Run / Hold +/-
press causes the SA to toggle between continuous sweep and
Hold.
——OFF
Causes the SA to Hold. Each Run / Hold +/- or
press causes the SA to make ONE sweep, then Hold. The Hold
annotation changes to an arrow --> while the sweep occurs.
—Single
Automatically sets Continuous OFF and causes the SA to make
ONE sweep, then Hold.
—Restart
Resets the Average counter to 1. The Average counter also is
used with MaxHold and MinHold Trace States. Learn more in
"Average Count" on page
Points [current setting]
101 | 201 | 401 | 601 | 801 | 1001

# Points

to enter a trace points value manually (Maximum = 20001).
All SA settings are applied to all displayed traces.
(IQA and SA Mode)" on page
or Run / Hold +/- button is pressed. Use Hold /
143.
194.
Single
See also
"Marker Trace
141