Calibration for NA, CAT, and VVM Modes

Mechanical Cal

Mechanical Calibration is performed using discrete standards from a Cal Kit.
Several Cal Kit definitions are built into the FieldFox. To learn about Cal Kit
definitions, refer to the Application Note, "Specifying Calibration Standards
and Kits for Keysight Vector Network Analyzers," available online at
http://literature.cdn.keysight.com/litweb/pdf/5989-4840EN.pdf
Visit
supported Cal Kits. Also at this website, download Data-Link software that
allows you to edit Cal Kit definitions or add a new Cal Kit.
Mechanical Cals are extremely accurate when performed using the correct Cal
Kits with standards that are clean and in good repair, and when using correct
connection procedures.
How to perform a Mechanical Cal
— Disconnect the DUT from the FieldFox.
— If a jumper cable or adapter is required to connect the DUT to the FieldFox,
— In NA, CAT, or VVM Mode, press Cal 5.
— Then
This page summarizes the Mechanical Cal to be performed and allows you to
make changes. For best results, review the screen and make changes in the
following order:
Keysight N9938-90003 User's Guide
www.keysight.com/find/fieldfoxsupport
then connect those components to the FieldFox connectors. The effects of
those components will be measured and removed during the calibration,
and only the effects of the DUT will be displayed in the measurement
results. These should be high-quality components!