Security Level - Keysight B Series User Manual

Fieldfox analyzers
Table of Contents

Advertisement

System Settings
System Configuration
The voltage is output through an SMB (m) output connector located on the
left-side panel. See the connector in
An optional SMB (f) to BNC (m) bias-tee power cable is available with Keysight
part number: N9910X-713.
The following status line is provided near the top of the FieldFox screen.
Figure 9-9
Voltage Source Status Line
1. Voltage source state: ON, OFF, TRIPPED.
2. Voltage setting from 1.00 V to 32.00 V with.01 V resolution.
3. Current: The amount of current (in Amperes) being drawn from the source.
4. Max Current: A conservative estimate of maximum current possible given
How to make Voltage Source settings
— Press System 7
— Then
— Then
— Then
— Then Voltage x.xx V Using

Security Level

For security reasons, you can prevent frequency information from appearing on
the FieldFox while in SA, NA, or CAT modes.
How to set Security Level
— Press System 7
— Then
252
the present voltage setting. The actual 'trip' level may be slightly different.
System
Configuration.
Voltage
Source.
Source ON OFF
to enable and disable the voltage source output.
keypad, enter a value between 0.001 and 32 volts.
Clear Overload
If the voltage source becomes 'TRIPPED' press to re-enable
the voltage source output. Then set
Status Line ON OFF
Turn ON and OFF the voltage source status line that
appears near the top of the screen. The status line is always ON while
making Voltage Source settings.
System Configuration
"Left Side Panel" on page
arrows, the rotary knob, or numeric
Source
ON.
Keysight N9938-90003 User's Guide
32.