Calibration for NA, CAT, and VVM Modes
— Press
'Unknown THRU': When prompted: "Connect port 1 to port 2", then
connect port and port 2, using a coaxial cable or an appropriate thru
adapter. Do NOT use the internal ECal THRU, because the ECal thru
calibration model is incorrect for this calibration type.
'Flush THRU': When prompted: "Connect port 1 directly to port 2", then
you must connect the ports directly together at the calibration reference
plane with no adapter or other cabling.
Learn more in

Simple Response Cals

Simple Response Cals are used to quickly calibrate the magnitude and phase
of a measurement using any Open, Short, or Thru component. These may be
calibration standards, but because a Cal Kit is not selected, they are not
modeled. Measurement accuracy is generally low. Use a Simple Response Cal
to make quick measurements when using a jumper cable to connect the DUT
to the FieldFox. Otherwise, CalReady is usually more accurate.
You can perform a Simple Response Cal for either: S11, or S22, or S21 AND
S12.
When prompted, choose a standard based on the displayed measurements to
be calibrated. For example, to calibrate S11, connect either an OPEN or
SHORT to the port 1 reference plane.
When Simple Response Cals are performed, the source match and reflection
tracking terms from CalReady are updated by the measured Short or Open that
is used during the calibration.
1-port response cals are also available from the Mechanical Cal menu. Learn
more in
Normalize uses a THRU standard or cable between port 1 and port 2 to cal an
S21 and S12 Transmission measurement (NA Mode) and a 2-port Insertion
Loss measurement (CAT Mode). In VVM Mode, this is performed using Zero.
When a Normalization is performed, the forward and reverse transmission
tracking terms from CalReady are updated to account for the THRU that is
used during the normalization process.
How to perform a Simple Response Cal
— Select the measurements to be calibrated. See the relevant Mode (NA, CAT,
96
Start
Calibration. When prompted, verify the ECal module
connection, then press Measure. The standards within the ECal module are
automatically connected and measured.
"Calibration Type" on page
"Calibration Type" on page
or VVM) for measurement selections.
98.
98.
Keysight N9938-90003 User's Guide