Calibrating Metropolix Inputs
You can calibrate all three of Metropolix' variable inputs ( X , Y , and Z ) using the same technique.To
Calibrate Tête's inputs:
TIP : If you make a mistake at any point during the following calibration procedure, you can scroll to
the bottom of the Calibration menu, select Reload All , and press the encoder. This will reload the
current, previously saved values back into Metropolix. Once you enact a Save All , this is no longer
possible, since your edited values will overwrite the previous calibration.
To Calibrate Metropolix' Inputs:
1.
Enter Calibration Mode, as discussed above.
2.
Prior to calibrating Metropolix' inputs, you must calibrate its outputs using the technique
discussed above.
Calibrated outputs are a necessity, since they will be used as the reference source for
calibrating Metropolix' inputs.
3.
Patch Metropolix' calibrated OUT A jack into Metropolix' X input jack
Although this example uses OUT A as a reference source, you can use any of the four
calibrated outputs: T RK 1 PITCH , TRK 2 PITCH , OUT A or OUT B .
NOTE: The positions of the X, Y, and Z attenuverters are ignored while calibrating.
4.
Turn the encoder to highlight ' Aux X: [number] / [number]' , then press the encoder.
Input X is automatically calibrated upon pressing the encoder.
5.
Move one end of the patch cable to the Y input; turn the encoder to highlight ' Aux Y: [number] /
[number]' , then press the encoder.
Input Y is automatically calibrated upon pressing the encoder.
6.
Move one end of the patch cable to the Z input; turn the encoder to highlight ' Aux Z: [number] /
[number]' , then press the encoder.
Input Z is automatically calibrated upon pressing the encoder.
To Save the Calibration:
1.
Turn the encoder to select 'Save All' , then press the encoder to save your calibration.