GE Voluson P8 BT16 Basic Service Manual page 489

Table of Contents

Advertisement

GE
DRAFT
DIRECTION 5723243, R
Section 8-44
Replacement of the HDD in DRFM
8-44-1
Manpower
One person, 30 minutes
8-44-2
Tools
Philips screwdriver
8-44-3
Preparations
1.) Power Off/Shutdown the system as described in
2.) Remove the REAR CABINET BEZEL ASSY BT16 and disconnect cables as described in
"Removal Procedure" 2), 3), 4) and 7) on page
3.) Remove the NEST EMI BRKT BT16 as described in
4.) Remove the DRFM as described in
8-44-4
Removal Procedure
1.) Unscrew 4 screws and pull the HDD out of the DRFM.
8-44-5
Installation Procedure
1.) Install the new parts in the reverse order of removal.
2.) Set up the new hard disk as described in
3.) Perform:
6
DRAFT (M
EVISION
Figure 8-201 Pulling the HDD
FRU8-32: Replacement of the HDD in DRFM - Functional Tests.
Chapter 8 - Replacement Procedures
23, 2018)
AY
4-3-2 on page
8-93.
8-56-4 "Removal Procedure"
8-42-3 on page
8-130.
5-17-5-3-2 on page
V
™ P8 / V
OLUSON
OLUSON
B
S
ASIC
ERVICE
4-4.
5-72.
™ P6
M
ANUAL
8-28-4
8-141