DIRECTION 5723243, R
5-11-1-3
Power Up Sequence Description
The Power Up Sequence can be divided in the following steps:
Switch Circuit Breaker in DPS to ON position
Press the ON button on the Operator Panel
DRFM power-up
5-11-1-3-1
Power Up Sequence Description
1.) DRFM, powered by 5Vstb, detects contact of Power-On button event.
2.) DRFM sends PSON_N low to DPS
3.) DPS powers up +24V, +12V,
4.) DPS provides LV_OK as soon as +12V is within specification
5.) DPS provides PC voltages from +12V, +5V.
6.) DRFM distribute enable signals for +3.3V, +2.5V, +1.2V, +1.1 to itself and DBM64.
7.) DRFM provides PS_DONE output signal to DPS.
8.) All Supply voltages are switched on.
6
DRAFT (M
EVISION
Figure 5-21 Power Up Sequence
Chapter 5 - Components and Functions (Theory)
23, 2018)
AY
±
5V
V
™ P8 / V
OLUSON
OLUSON
B
S
M
ASIC
ERVICE
™ P6
ANUAL
5-43