GE Voluson P8 BT16 Basic Service Manual page 353

Table of Contents

Advertisement

GE
DRAFT
DIRECTION 5723243, R
8-2-1-1
System Functional Checks
Table 8-2
Mode
2D Mode
Quality
Receiver
Frequency
M Mode
Volume
Mode
Triplex
Mode
NOTE:
* The use of test phantoms is only recommended if required by your facility 's (customer's)QA program.
6
DRAFT (M
EVISION
System Function
Connect an abdominal probe, press the PROBE key, select the
"Abdomen" - Application and start the "Default" -Program.
Record a 2D image of the Liver.
If there is no abdominal probe, record a 2D image the Thyroid
using a small parts probeand corresponding program.
Use the FREQUENCY control to switch the Receiver Frequency
range (penet./norm/resol.).
Start the "Abdomen" -Program, adjust the M Cursor (vessel) and
activate the M Mode. Adjust the SPEED key to change the M
Mode sweep speed. After FREEZE, move the TRACKBALL to
recall the stored sequence.
Start the Volume acquisition using a Real Time 4D probe.
Start the "Abdomen" -Program and switch on the CFM- and the
PW Mode. Adjust the Doppler Cursor and press the right trackball
key to activate the Triplex-Mode.
Chapter 8 - Replacement Procedures
23, 2018)
AY
Task
V
™ P8 / V
OLUSON
OLUSON
B
S
ASIC
ERVICE
Expected Result
regular and homogenous 2D image
no disturbances in the 2D image during
changing the Receiver Frequency
the M Cursor agree with Vessel Cine
loop is displayed
continuous Volume Acquisition (without
any "jumps") and smooth echo shape
with clear defined image edge
no disturbances in the 2D/Color and
the Doppler image
™ P6
M
ANUAL
8-5