GE
RAFT
DIRECTION 5723243, R
Section 8-4
Image Settings Only - Loading Procedure
8-4-1
Introduction
The Image Settings contain:
•
Application Settings
•
2D Factory and 2D user Presets
•
3D/4D Factory and 3D/4D User Presets
•
Annotation Presets
•
Scan Assistant Configuration
•
Measure configuration
•
Biopsy Lines
8-4-2
Loading Procedure
see:
4-5-2 "Load Small Backup (Scan Settings)" on page 4-13
Section 8-5
Full Backup (Full System Configuration) - Loading Procedure
8-5-1
Introduction
The Full Backup contains following data:
•
User Settings (databases and files containing User Programs, 2D/3D/4D Programs, Auto Text
entries, gray curves and complete System settings such as language, time/date format, etc.)
•
Measure Configuration (user specific measure setup settings)
Patient Archive (database containing patient demographic data and measurements) - no images
•
•
Options (Permanent Key that is specific for enabled software options and Demo Key)
•
Image Transfer Configuration (DICOM settings e.g., DICOM servers, AE Title, Station Name, etc.)
•
Network Configuration (network settings including the computer name)
•
Service Platform (state of the Service Software)
8-5-2
Loading Procedure
see:
4-5-4 "Load Full System Configuration (Full Backup)" on page 4-18
Section 8-6
Image Archive - Loading Procedure
8-6-1
Introduction
A backup of the Image Archive contains the Patient Archive (database containing patient demographic
data and measurements) + images.
8-6-2
Loading Procedure
see:
4-5-6-2 "Load Image Archive" on page 4-23
8-22
6
DRAFT (M
EVISION
Section 8-4 - Image Settings Only - Loading Procedure
23, 2018)
AY
V
™ P8 / V
OLUSON
OLUSON
B
S
M
ASIC
ERVICE
™ P6
ANUAL