DIRECTION 5723243, R
7-10-3
Trackball - Impaired Sensitivity
The trackball has low sensitivity!
Set up the appropriate trackball
response time
(via [Trackball Speed] button in the
System Setup - User Settings page).
Still low sensitive...
Remove the trackball and clean it from dust.
Remove the trackball and clean it from dust.
Mount the outer fixing ring by turning it clockwise.
Return to normal scanning
7-46
Section 7-10 - Troubleshooting Trees, Instructions and Tech Tips
6
DRAFT (M
EVISION
Is it difficult to roll the
trackball?
YES
Close
Figure 7-40 Trackball - Impaired Sensitivity
23, 2018)
AY
Contact GE service representative
NO
to replace the trackball module.