GE Voluson P8 BT16 Basic Service Manual page 265

Table of Contents

Advertisement

GE
DRAFT
DIRECTION 5723243, R
5-15-1-3
Rollback
This function offers the possibility to simply restore the previously used system configuration (rollback),
which was stored on R:\ during "FMI from DVD".
1.) Turn system OFF and then back ON.
2.) As soon as the "Boot Screen" appears (see:
[ Arrow down] key on the keyboard until the Rollback item is highlighted, then press ENTER.
After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item).
CAUTION
If you missed step 2, retry again with CTRL + ALT + DEL.
3.) When the following WARNING message appears, press the Arrow left] button to highlight OK
and then press ENTER.
select [OK]
After performing the rollback, the system reboots.
The Ultrasound Application is started and finally the 2D screen is displayed on the monitor.
6
DRAFT (M
EVISION
Figure 5-32 Warning message
Chapter 5 - Components and Functions (Theory)
23, 2018)
AY
Figure 5-31 on page
V
™ P8 / V
OLUSON
OLUSON
B
S
M
ASIC
ERVICE
5-56), press the PG DN
™ P6
ANUAL
5-57