5-2-1-4
Pulsed (PW) Doppler
PW Doppler processing is one of two spectral Doppler modalities. In spectral Doppler, blood flow is
presented as a scrolling display, with flow velocity on the Y-axis and time on the X-axis. The presence
of spectral broadening indicates turbulent flow, while the absence of spectral broadening indicates
laminar flow. PW Doppler provides real time spectral analysis of pulsed Doppler signals. This
information describes the Doppler shifted signal from the moving reflectors in the sample volume. PW
Doppler can be used alone but is normally used in conjunction with a 2D image with an M-line and
sample volume marker superimposed on the 2-D image indicating the position of the Doppler sample
volume. The sample volume size and location are specified by the operator. Sample volume can be
overlaid by a flow direction cursor which is aligned, by the operator, with the direction of flow in the
vessel, thus determining the Doppler angle. This allows the spectral display to be calibrated in flow
velocity (m/sec.) as well as frequency (Hz). PW Doppler also provides the capability of performing
spectral analysis at a selectable depth and sample volume size. PW Doppler can be used in
combination with 2D and Color Flow modes.
5-2-1-5
3D Imaging
The Voluson™ P8 / Voluson™ P6 Ultrasound System will be used to acquire multiple, sequential 2D
images which can be combined to reconstruct a three dimensional image. These 3D images are useful
in visualizing three-dimensional structures, and in understanding the spatial or temporal relationships
between the images in the 2D sequence. The 3D image is presented using standard visualization
techniques, such as surface or volume rendering.
5-2-1-6
3D Data Collection and Reconstruction
2D gray scale images including Color Flow or Power Doppler information may be reconstructed.
The acquisition of volume data sets is performed by sweeping 2D-scans with special transducers
(called 3D-transducers) designed for the 2D-scans and the 3D-sweep.
2D ultrasound imaging modes are used to view a two dimensional cross-sections of parts of the body.
For example in 2D gray scale imaging, a 2 dimensional cross-section of a 3-dimensional soft-tissue
structure such as the heart is displayed in real time. Typically, the user of an ultrasound machine
manipulates the position and orientation of this 2D cross-section in real time during an ultrasound exam.
By changing the position of the cross-section, a variety of views of the underlying structure are obtained,
and these views can be used to understand a 3-dimensional structure in the body.
To complete survey a 3-dimensional structure in the body, it is necessary to collect 2D images which
span a volume containing the structure. One way is to sweep the imaging cross-section by translating
it in a direction perpendicular to the cross-section. Another example method is to rotate the cross
section about a line contained in the cross section. The Voluson™ P8 / Voluson™ P6 Ultrasound
System uses the automated so called C-Scan for the motion perpendicular to automated B-scan. Once
a representative set of 2D cross-sections are obtained, standard reconstruction techniques can be used
to construct other 2D cross-sections, or to view the collection of the cross-sections as a 3D images.
5-2-1-7
3D Image Presentation
Several techniques can be used to aid the human observer in understanding the resulting 2D image as
a representation of a three-dimensional object. One is to rotate the volume of data, and present the
resulting sequence of 2D projections to the observer. The changing direction of observation helps the
observer to separate the features in the volume according to their distance from the observer.
5-8
6
DRAFT (M
EVISION
Section 5-2 - General Information
23, 2018)
AY
V
™ P8 / V
OLUSON
OLUSON
B
S
M
ASIC
ERVICE
™ P6
ANUAL

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson p6 bt165670023567002456699995670021

Table of Contents