GE Voluson P8 BT16 Basic Service Manual page 185

Table of Contents

Advertisement

GE
DRAFT
DIRECTION 5723243, R
4-5-6
Archiving Images
It is recommended to Save Image Archive
CAUTION
System Configuration
!! CAUTION:
4-5-6-1
Save Image Archive
NOTE:
A backup of the Image Archive always contains the Patient Archive (database containing patient
demographic data and measurements) + images of the selected exams.
1.) Press the F2 key on the control panel.
2.) In the "Utilities" menu select SYSTEM SETUP to invoke the setup desktop on the screen.
3.) On the left side of the screen select BACKUP and then click the
4.) Click the SAVE button.
6
DRAFT (M
EVISION
(Section 4-5-3 on page
Figure 4-23 System Setup - Backup - IMAGE ARCHIVE page
Figure 4-24 Image Archive Save - preparations
Chapter 4 - Functional Checks
23, 2018)
AY
(Section 4-5-6-1 on page
4-16) once a week.
V
™ P8 / V
OLUSON
OLUSON
B
S
ASIC
ERVICE
4-21) and Backup the Full
IMAGE ARCHIVE
tab.
[Load] button
[Save] button
[Advanced] button
choose archiving
span time
check mark if desired
click [Next] button
™ P6
M
ANUAL
4-21