DIRECTION 5723243, R
5-14-3-14
Probe EEPROM
NOTE:
It is recommended to connect and re-program one probe after the other.
Do not assign wrong coding (e.g., programming a RAB to a RIC probe). An inaccurate probe
CAUTION
programming will mostly likely damage the probe and/or system.
1.) Click the Probe EEPROM button on the "Service Tools" page (see
Tools page" on page 5-50
2.) Check and/or choose correct probe name from the pull-down menu (1).
3.) Check and/or enter probes Serial Number in the appropriate [SNO] filed(2).
4.) Select "Today" from the Start date pull-down menu(3).
5.) Click the OK ( Global EEPROM) button(4).
NOTE:
DO NOT disconnect the probe during the re-programming process!
Probe ID and coding is assigned to the probe that is connected to the selected prot.
6.) Confirm with OK.
6
DRAFT (M
EVISION
Figure 5-30 Probe EEPROM
Chapter 5 - Components and Functions (Theory)
23, 2018)
AY
V
™ P8 / V
OLUSON
OLUSON
B
S
M
ASIC
ERVICE
Section Figure 5-25 "Service
™ P6
ANUAL
5-55