GE Voluson P8 BT16 Basic Service Manual page 259

Table of Contents

Advertisement

GE
DRAFT
DIRECTION 5723243, R
5-14-3-1
Auto Tester
Autotester is a log function of customer activities. It records all user actions (scanning, entries,
performing Calculations, review of Patient Reports, etc....). It is possible to safe (record) as file on HDD.
But also export to DVD/CD can be done to allow replay of the records on other units.
NOTICE
For intermittent problems this file can be requested from the Service Engineer or customer.
It is possible to burn this file on DVD/CD+R/RW.
!! NOTICE:
Operation see:
5-14-3-2
Update
5-14-3-2-1
FMI from DVD
By means of the FMI FROM DVD button, the Systems C:\ image is partly or completely updated.
The System Software parts to be upgraded depend on contents of the used System DVD.
The first "Boot Device" in BIOS has to be Hard Disk Drive. To change BIOS Boot Priority Order
PRECONDITION:
setting, see
NOTE:
During "FMI from DVD" the used system configuration (incl. Full Backup) will be stored on R:\.
If required, the previously used System configuration (before FMI from DVD was performed) can
be restored by activating the "Rollback" function. Operation see:
The Linux script uses a table, which describes the procedure to upgrade the current master image by
the new one on the DVD. This table is sored as CSV text on the System DVD. It logs the development
history of the master images via a successive Svn. nr. At the same time the steps necessary for
upgrading an old image with a new one are defined.
# svn nr
mandatory buegeln
application
MS patches
peripherals
insite
settings
EUM
image
last mandatory FMI svn nr
6
DRAFT (M
EVISION
Section 7-6 "How to use the Auto Tester program" on page
Figure 5-26 development history table (example)
Successive version number of the master image.
B means that the systems C:\partition will be completely replaced by the image on the
DVD
name/number of the Ultrasound Application Sotware
name of the MS patch(es)
name of the Peripherals driver installation program
directory name of Insite service platform updates
directory name of the new User Settings ( Application Settings)
Electronic User Manual(EUM) files
version of image when "mandatory buegeln" = B
version number of the last mandatory FMI
Chapter 5 - Components and Functions (Theory)
23, 2018)
AY
V
™ P8 / V
OLUSON
OLUSON
B
S
M
ASIC
ERVICE
7-29.
Section 5-15-1-3 on page
™ P6
ANUAL
5-57.
5-51

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson p6 bt165670023567002456699995670021

Table of Contents