Network Troubleshooting
7-10-7-1
No Connection to the Network at All
Gray
= Cable disconnected or no network signal on a connected cable.
Green
= Cable connected to a network. Does not imply proper network settings.
1.) Check that the network cable between the VVoluson™ P8 / Voluson™ P6 system and the wall
network is connected and well seated in both ends. (Use a network cable that is known to be OK.)
2.) Connect a network cable between the system and a PC by either using a hub or a cross-over cable.
Try to ping from system to IP address on PC. If OK, hardware connection inside the system is OK.
7-10-7-2
GE remote service connection
The customer gets visual information about GE remote service status (shown in the status area on the
right bottom of the screen).
Gray = Idle State
Yellow = Access
pending
Red = Disrupted State
Gray + envelope
6
DRAFT (M
EVISION
Remote access is inactive.
Remote Access connection is active, but Disruptive Mode and Virtual Console
Observation (VCO) is not yet enabled.
Remote Access is active. All processes [UL_VNC and UL_CSD] are active. In this
state the Voluson P-Series system should NOT be used clinically.
New Software Update Package is available for download and installation. Operation
see
Section 8-2-4 "Software Update Package - Download/Installation Procedure" on
page
8-13.
Chapter 7 - Diagnostics/Troubleshooting