7.8 Updating software

7.8.1 Performing a software update
Requirement
1. If necessary: Download software from the Login area of the
2. If the software is available as a ZIP file: Unzip the software.
3. Place the file in the SD card's root directory.
4. Insert SD card
5. Select the Software update menu item.
6. Select Software update.
The device is connected to line power.
The operator menu is activated
operator menu", page
WEINMANN Emergency website to the SD card.
The folder contains the software in a file named
WM35756-x.x.hex.
When doing so: The software update file must not be in
a sub-folder.
(see "7.7.1 Inserting an SD card", page
(see "7.1 Activating the
67).