Pump Repair
32

Removing the Run Button

32.1
Remove the control PCB assembly from the pump (Refer to Page 27, Section 22).
32.2
Remove the button return spring (Fig 48, Item 60) and the Run button (Fig 48, Item 50) from the
top case assembly (Fig 48, Item 20).
33

Installing the Run Button

33.1
Install the Run button (Fig 48, Item 50) into the cutout in the top case assembly
(Fig 48, Item 20).
33.2
Install the button return spring (Fig 48, Item 60) onto the Run button.
33.3
Install the control PCB assembly into the top case assembly, and then re-assemble the pump
(Refer to Page 29, Section 25).
34

Replacing the Top Case Assembly

The top case assembly (Fig 48, Item 20) comprises the top case moulding and the membrane
label.
34.1
Remove the control PCB assembly from the top case assembly (Refer to Page 27, Section 22).
34.2
Remove the Run button (Refer to Page 40, Section 32).
34.3
Install the Run button into the new top case assembly (Refer to Page 40, Section 33).
34.4
Install the control PCB assembly into the top case assembly, and then re-assemble the pump
(Refer to Page 29, Section 25).
Chap 5
Page 40
Flowtron Universal Service Manual
SER0009