Flowtron Universal Service Manual
Make sure the mains/power
cordset connector is pushed
securely into the pump
Note: Insert the end of the test probe through the test hole in the rear case.
Approximate dimensions of test probe: Minimum length = 100 mm
Figure 30 - Inserting the Test Probe through the Rear Case
Figure 31 - Earth/Ground Test Area on Control PCB Assembly
Issue 2
Sep 2009
Maximum diameter = 2 mm
SER0009
Testing
Test Probe
Screw
Case IEC Cover
Earth/Ground
Access Point
Rear Case of Pump
Mains/Power Cordset
Control PCB Assembly
Front Case of Pump
IEC Connector
Earth/Ground Test Pad
(inside dotted line)
Compressor Assembly
Chap 4
Page 13