5

Replacing the Inlet Filter Felt

5.1
Remove the case IEC cover (Refer to Page 6, Section 3).
5.2
Remove the inlet filter felt (Fig 33, Item 20) from the back of the case IEC cover
(Fig 33, Item 30), and discard.
5.3
Push the new inlet filter felt into the back of the case IEC cover.
5.4
Install the case IEC cover (Refer to Page 6, Section 4).
6

Removing the Mains/Power Cordset

6.1
Remove the case IEC cover (Refer to Page 6, Section 3).
6.2
Pull the mains/power IEC cordset connector (Fig 33, Item 10) out of the IEC connector inside
the pump.
7

Installing the Mains/Power Cordset

CAUTION: Make sure the replacement mains/power cordset (Fig 33, Item 10) is the
correct type and rating, or the electrical safety of the pump may be impaired.
7.1
Carefully push the mains/power IEC cordset connector (Fig 33, Item 10) fully into the IEC
connector inside the pump. Place the cordset cable into the groove in the rear case.
Note: Make sure the cordset cable is not pulled tight when it is installed into the rear case.
There should be some slack in the cordset cable under the case IEC cover.
7.2
Install the case IEC cover (Refer to Page 6, Section 4).