Checking The Version Of The Pump Software; Downloading The V3.01 Software Files To The Pump - Arjohuntleigh FLOWTRON UNIVERSAL Service Manual

Hide thumbs Also See for FLOWTRON UNIVERSAL:
Table of Contents

Advertisement

Flowtron Universal Service Manual
9

Checking the Version of the Pump Software

9.1
Look at the LCD screen, and make sure the pump is in Standby (refer to Figure 85).
9.2
On the Communicator software, click on the Command List tab (refer to Figure 84).
9.3
Scroll through the menu of Flowtron Universal Commands on the Communicator software to
the section heading VERSIONS.
9.4
Send the command 801 - Read_Software_Version to the pump.
9.5
Check the Receive part of the dialog box:
9.5.1
There should be 801 = 3.01 in the Receive part of the dialog box, where 3.01 is the
correct version of the software downloaded to the pump. This is the end of the
software upgrade. Continue the test at Para 9.6 below.
9.5.2
If the command is unsuccessful or the software version is NOT 3.01, then:
9.5.2.1
9.5.2.2
9.6
Disconnect the pump from the mains/power supply.
9.7
Remove the Serial Communications Pack from the pump (refer to Chapter 6 "Updating the
software", Page 9, "Connecting the Pump to a Laptop / Desktop PC").
10

Downloading the V3.01 Software Files to the Pump

10.1
Load the V3.01 Flowtron Universal pump software onto the PC, as follows:
10.1.1
The software for the Flowtron Universal pump is contained in the following three files
which must be downloaded to the pump using the Communicator software:
FlowtronUniversal_V3-01.H00
FlowtronUniversal_V3-01.H01
FlowtronUniversal_V3-01.H02
10.1.2
These three software files will normally be supplied on either of the following media:
CD ROM.
File attachments on an email.
10.1.3
If the three software files are supplied on a CD ROM, load the CD ROM onto the PC.
10.1.4
If the three software files are supplied as attachments on an email, do the following:
10.1.4.1
10.1.4.2
10.1.4.3
10.1.4.4
10.2
Initialise the pump (refer to Page 15, Section 8).
10.3
Look at the LCD screen, and make sure the pump is in Standby (refer to Figure 85).
10.4
On the Communicator software, click on the Bank Loader tab (refer to Figure 84).
10.5
Click the middle button at the top right of the dialog box to maximise the screen to view the
Tx Rx box at the bottom of the dialog box.
Note:
At this stage, ignore any text in the Tx Rx box.
10.6
On the left side of the dialog box, under File Names, select the three pump software files, as
follows (refer to Figure 84):
10.6.1
Click in the boxes on the right of Bank 0, Bank 1 and Bank 2 to enable them. The text
will become visible and there will be a tick (
Issue 2
Sep 2009
The software upgrade has failed and the procedure must be repeated.
The pump is faulty and must be rejected.
Open the email.
Click on File, followed by Save Attachments.
Make sure all three files are selected, and click OK.
Browse in the Save Attachment dialog box for a suitable location to save
the two files, and click Save.
SER0009
Updating the software
) in each box.
Chap 6
Page 17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents