Testing
9

Recalibrating the Pump Hose Set Assemblies

9.1
Initialise the pump (Refer to Page 1, Section 1).
9.2
Make sure that no garments are connected to either pump hose set assembly.
9.3
Press the "hidden" Service button on the membrane panel to access the service screen on the
LCD display (Refer to Page 5, Section 4).
9.4
Using the Down and Up buttons (Refer to Page 5, Section 5), scroll down the list of options to
Calibrate RFID, and select it using the
9.5
A new list of options will appear. Scroll down the list of options to Save as No Garment, and
select it using the
9.6
Scroll down the list of options to Exit, and select it using the
9.7
If at any time the LCD display shows Pump Fault - RFID, then the hose set electrical
connections are not fitted correctly at the pump end of the hose set assemblies. Check and
rectify the electrical connections, or replace the hose set assemblies.
Chap 4
Page 12
button (Refer to Figure 28).
Figure 28 - LCD Display - Calibrate RFID
button (Refer to Figure 29).
Figure 29 - LCD Display - Save as No Garment
SER0009
Flowtron Universal Service Manual
button, to end the test.
Issue 2
Sep 2009