Pump Repair
8.4
Carefully lift the elbow retaining cover and elbow connector out of the pump case to reveal the
cable connection between the elbow retaining cover and the pump.
8.5
Disconnect the hose set cable connector from the 6-way connector on the control board
assembly in the pump (Refer to Figure 37), and remove the hose set assembly from the pump.
Elbow Retaining Cover
in Secured Position
Elbow
Connector
9

Installing the Hose Set Assembly

9.1
If you want to re-fit the same hose set assembly which was removed from the pump, then
inspect the screw cover on the elbow retaining cover on the hose set assembly (Refer to
Figure 35). If the screw cover or hinge is damaged or split, then replace the hose set assembly.
Note: It is not necessary to replace both hose set assemblies if only one is damaged.
9.2
Put the hose set assembly (Fig 34, Item 10) by the pump, with the elbow retaining cover
adjacent to the case cutout in the top of the pump.
9.3
Plug the hose set cable connector into the 6-way connector on the control board in the pump.
Note: Make sure the hose set cable connector is pushed fully into the control board connector.
9.4
Carefully insert the excess hose set cable inside the pump case, between the control board
connector and the threaded pillar (Refer to Figure 37).
Note: Do not insert the excess hose set cable the other side of the control board connector or it
might get damaged by the lugs on the elbow retaining cover.
Note: Make sure the cable is always clear of the pneumatic outlet in the pump.
9.5
Install the elbow retaining cover fully into the cutout in the pump case.
Note: Make sure the two lugs on the elbow retaining cover are aligned with the correct slots in
the cutout in the pump case (Refer to Figure 37).
Chap 5
Page 10
Screw Cover
Elbow
Retaining
Cover
Tube
Figure 36 - Releasing the Hose Set Connector
SER0009
Flowtron Universal Service Manual
20 degrees
Elbow Retaining Cover
in Free Position
Turn Tube and
Elbow Connector
counterclockwise
Issue 2
Sep 2009