System Operation - Arjohuntleigh FLOWTRON UNIVERSAL Service Manual

Hide thumbs Also See for FLOWTRON UNIVERSAL:
Table of Contents

Advertisement

Introduction
4

System Operation

4.1
Plug the pump into an electrical outlet, and switch on.
4.2
The pump will first go through a self-test routine (Refer to Figure 2).
4.3
At the end of the self-test routine, the pump will "beep" and go into Standby mode, where it will
remain until required for use (Refer to Figure 3).
4.4
Connect the garments to the pump hose set assemblies and make sure the connectors click into
place.
4.5
Check that the connection and garment type are confirmed correctly on the display. Figure 4
shows two single-chamber calf garments connected to the pump.
4.6
Typical garment options and default therapy parameters are detailed in Table 1.
Table 1 - Garment Options and Default Therapy Parameters
Garment Type and
LCD Display Graphic
Single-chamber,
Calf and Thigh/Calf
Chap 1
Page 4
Figure 2 - LCD Display - Running Diagnostic
Figure 3 - LCD Display - Standby
Figure 4 - LCD Display - Garments Connected
Inflation Pressure
Foot
SER0009
Flowtron Universal Service Manual
Inflation Hold
(mmHg)
(seconds)
130
40
Cycle
(seconds)
3
30
12.5
60
Issue 2
Sep 2009