Pump Repair
3

Removing the Case IEC Cover

WARNING: BEFORE REMOVING THE CASE IEC COVER, MAKE SURE THE PUMP UNIT
HAS BEEN ISOLATED FROM THE MAINS/POWER SUPPLY BY REMOVING
THE MAINS/POWER PLUG FROM THE WALL SOCKET.
3.1
Lay the pump face down on a flat surface.
3.2
Remove the two screws (Fig 33, Item 40), the case IEC cover (Fig 33, Item 30) and the inlet
filter felt (Fig 33, Item 20) out of the rear case.
Note: The case IEC cover is a tight fit in the rear case. Removal of this cover is made easier if
the two screws are removed completely from the cover, and then the mains/power
cordset (Fig 33, Item 10) is gently moved up and down while the cover is removed.
4

Installing the Case IEC Cover

4.1
Make sure the mains/power IEC cordset connector (Fig 33, Item 10) is pushed fully into the IEC
connector inside the pump. Place the cordset cable into the groove in the rear case.
Note: Make sure the cordset cable is not pulled tight when it is installed into the rear case.
There should be some slack in the cordset cable under the case IEC cover.
4.2
Push the case IEC cover (Fig 33, Item 30) and inlet filter felt (Fig 33, Item 20) into the slot in the
rear case.
4.3
Install the two screws (Fig 33, Item 40) to secure the case IEC cover.