Pump Repair
39.4
At the other end of the compressor assembly, do the following (Refer to Figure 60):
39.4.1
Hold the compressor assembly, and push the mounting bracket in towards the centre of
the compressor assembly.
39.4.2
At the same time, carefully push the grooved part of the remaining two AV mounts into
the slots in the mounting bracket.
39.5
Install the two compressor bump stops (Fig 59, Item 40) through the compressor bracket
grommets (Fig 59, Item 30), and over the spigots in the compressor assembly.
39.6
Install the two screws (Fig 59, Item 50) to secure the compressor bump stops.
39.7
Check that the two wire assemblies (long and short) are correctly connected to the compressor
assembly. Refer to Figure 61 for the correct wire colours and compressor terminal connections.
Note: The two wire assemblies are part of the compressor assembly.
39.8
Install the compressor assembly and mounting bracket into the pump (Refer to
Page 43, Section 37).
Short Wire Assembly
Wire Colour Key:
White
Black
Wire Assembly Connections:
Short (Item 20): White wires: 6 and 8
Black wires: 5 and 7
Chap 5
Page 46
Compressor Assembly
5
6
Long Wire Assembly
Long (Item 30): White Wires: 2 and 4
Figure 61 - Compressor Assembly
SER0009
Flowtron Universal Service Manual
7
8
Compressor Terminal
2
1
Air Inlet Filter Holder
Black Wires: 1 and 3
Identification
4
3
Air Inlet Filter