Page 5
• Keep the pump away from sources of liquids and do not immerse in water. • Do not use the pump in the presence of uncontained flammable liquids or gasses. • Only the pump and garment/insert combination as indicated by ArjoHuntleigh should be used. The correct function of the product cannot be guaranteed if incorrect pump and garment combinations are used.
Page 6
ArjoHuntleigh distributor. Do NOT use unapproved accessories or attempt to modify, disassemble or otherwise misuse the Flowtron Universal system. Failure to observe this caution could result in injury, or in extreme cases, death. Environmental Protection Incorrect disposal of this equipment and its component parts, particularly batteries or other electrical components, may produce substances that are hazardous to the environment.
If you have any difficulties in setting-up or using the system, contact your local Flowtron Universal ArjoHuntleigh sales representative, listed at the end of this manual. Intended Use The intended use of this product is to prevent Deep Vein Thrombosis (DVT). The garments are single use only. It is not for use in the home healthcare environment.
2. Clinical Applications Indications The primary application of the Flowtron Universal system is for the prevention of Deep Vein Thrombosis (DVT), when combined with an individualised monitoring programme. These systems represent one aspect of a DVT strategy; if the patient's condition changes the overall therapy regimen should be reviewed by the prescribing clinician.
4. When used for DVT prevention, continuous external pneumatic compression is recommended until the patient is fully ambulatory. Uninterrupted use of the system is Flowtron Universal encouraged. 5. The system should be USED Flowtron Universal WITH CAUTION on patients with: •...
Page 10
• ArjoHuntleigh does not recommend the use of compression stockings with its system. If these are ordered by the physician, the clinician should...
If the operation or performance of the pump changes during use, Refer to “Troubleshooting” on page 18 of this IFU before calling a service engineer or contacting your local ArjoHuntleigh sales office. Run/Standby Button LCD Display and Function Buttons...
Garment Application Use the correct garment type as ordered by the patient’s physician. For detailed instructions on correct fitting and use, carefully read and follow the instruction provided with every garment set. Connect the garments to the pump hose set connectors and ensure the connectors click into place.
This example shows the pump running and the left garment inflated to 40 mmHg. Refer to “Battery Operation” on page 12 when operating the pump under battery power. TO STOP press and hold the green RUN button a second time. The green power indicators will go out and therapy will stop.
To Use Different It is possible to use different garment types on each limb Garments on Each (if prescribed by the patient’s physician). Limb Connect the garments to either tubing outlet and the correct therapy cycle will be administered. Check on the display panel to confirm the garment types connected.
If the physician requires different therapy settings, limited changes can be made through a service link. Contact your local ArjoHuntleigh Sales Representative for details.
5. Options Pressing the option button will bring up a multi-choice menu of options. The option can be set using a combination of the three function buttons on the display. To return to the previous menu at any time, highlight using the arrow buttons and press the ...
Page 17
2. Use the arrow buttons to highlight ‘Language’. 3. Press the button to confirm. 4. Use the arrow buttons to scroll through the available languages. 5. Press the button to confirm. 6. Press the button again to confirm. 7.
6. Battery Operation When used with the optional battery pack, the system can be run independently of the mains power supply. Typically, 6-8 hours of operation can be expected from a fully charged battery pack. From the front of the pump, engage the pack on the Connecting the Battery Pack runners and slide into place until a click is heard.
15), remove and store the battery pack. If the battery pack is to be put into long term storage it should be recharged at least once every three to six months. A faulty battery pack should be sent back to ArjoHuntleigh for recycling or correct disposal.
Checking the Status of the Battery Battery Label LED 1 LED 2 Jack Socket Polarity 80-100% Charge Battery Pack OK Battery Pack 30-80% Charge needs recycling <30% Charge Battery Pack Fault Press Reset / Status Latch 1. Remove the battery from the pump unit. To Check the Battery Status: 2.
Avoid charging the battery near source of heat or in direct sunlight. Caution Only ArjoHuntleigh-approved battery chargers must be used. Use of a non-approved charger may result in damage to the battery pack or ineffective charging and therefore voids the warranty.
Healthcare Facility or the country of use. If you are uncertain, you should seek advice from your local Infection Control Specialist. The Flowtron Universal system should be routinely decontaminated between patients and at regular intervals while in use; as is good practice for all reusable medical devices.
8. Routine Maintenance Flowtron Universal System The equipment has been designed to be maintenance- Maintenance free between service periods. will make available on request service Servicing ArjoHuntleigh manuals, component parts lists and other information necessary for trained personnel to repair ArjoHuntleigh the system.
9. Troubleshooting pump features an audible and visual alarm. If a problem Flowtron Universal occurs, the system will sense the fault and briefly flash a message on the display. If the same fault continues for 10 successive inflations (up to 10 minutes), the audible alarm will sound, the red alarm lights flash and a message will appear on the display until corrective action is completed.
10. Accessories pump should only be used with the following garments: Flowtron Universal CALF GARMENTS Order Code Type Calf Circumference DVT10 DVT10 Standard Calf Garment Up to 43 cm (17") DVT10S DVT10S Standard Calf Garment (Sterile) Up to 43 cm (17")
ENVIRONMENTAL INFORMATION Condition Temperature Range Relative Humidity Atmospheric Pressure Operating +10 °C to +40 °C 30% to 75% 700 hPa to 1060 hPa (+50 °F to +104 °F) (non-condensing) Storage and Transport +10 °C to +40 °C 20% to 95% 700 hPa to 1060 hPa (Long Term) (+50 °F to +104 °F)
Page 28
Guidance and manufacturer’s declaration - electromagnetic immunity The pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should assure that it is used in such an environment. Immunity Test IEC 60601 Test Compliance Electromagnetic environment - guidance Level...
Page 29
Recommended separation distances between portable and mobile RF communications equipment and the pump The pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the pump as recommended below, according to the maximum output power of the communications equipment.
Page 31
ÖSTERREICH Τηλ: 21 0724 36 68 ArjoHuntleigh GmbH Φάξ: 21 0721 55 53 POLSKA Dörrstrasse 85 ArjoHuntleigh Polska Sp. z o.o. AT-6020 INNSBRUCK ESPAÑA ul. Ks Piotra Wawrzyniaka 2 Tel: +43 (0) 512 204 160 0 ArjoHuntleigh Ibérica S.L. PL 62-052 KOMORNIKI (Poznan) Fax: +43 (0) 512 204 160 75 Ctra.
Page 32
We operate under the three brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh focuses on patient mobility and wound management solutions. GETINGE provides solutions for infection control within healthcare and contamination prevention within life sciences.