Testing
2.3.2
Figure 14 shows a foot garment and a single-chamber calf or thigh/calf garment
connected to the pump.
Figure 14 - Foot and Single-Chamber Calf or Thigh/Calf Garments Connected
2.3.3
Figure 15 shows a foot garment connected to one hose set assembly, and no garment
connected to the other.
2.3.4
If the LCD display shows either an incorrect garment graphic or a "no garment" graphic
when one is connected, then carry out the procedure for recalibrating the pump for a
new hose set assembly (Refer to Page 12, Section 9).
Table 3 - Garment Options and Default Therapy Parameters
Garment Type
Foot
Single-chamber,
Calf and Thigh/Calf
No Garment Connected
Chap 4
Page 2
Figure 15 - Only One Garment Connected
LCD Display
Inflation Pressure
Graphic
SER0009
Flowtron Universal Service Manual
Inflation Hold
(mmHg)
(seconds)
130
40
12.5
-
Cycle
(seconds)
3
30
60
-
-
Issue 2
Sep 2009