Arjohuntleigh FLOWTRON UNIVERSAL Service Manual page 10

Hide thumbs Also See for FLOWTRON UNIVERSAL:
Table of Contents

Advertisement

Introduction
The system comprises a pump that can be used in conjunction with a pair of ArjoHuntleigh single-patient-
use compression garments.
The pump provides intermittent cycles of compressed air to the two connected compression garments,
which are inflated alternately. The pump can be used with a number of different ArjoHuntleigh single-
patient-use compression garments, including:
Foot garments, e.g. FG100-200.
Single-chamber calf and thigh/calf garments, e.g. DVT10-40, 60.
The pump automatically adjusts the correct therapy profile for the garment type being inflated.
An optional battery pack is available which allows the pump to be used independently of the mains/power
supply. The battery pack slides onto the base of the pump, and recharges itself when the pump is
operating from the mains/power supply.
The pump has the following controls (Refer to Figure 1):
A Run/Standby push-button. This is used to start and stop the therapy applied to the patient.
An illuminated graphic LCD display. This indicates:
The type of garment(s) connected.
The pressure delivered to each garment.
Alarm indications.
A membrane label which contains the following four soft-function buttons:
A three-button keypad below the LCD display. These are used during the normal operation
of the pump.
A single "hidden" button above the LCD display. This is used during service and repair
operations only.
Red and green LED indicators on the front of the pump to show alarm and run status.
Automatic detection of the connected garment type.
LCD Display +
Membrane Label
Power/Alarm LED
Indicators
Chap 1
Page 2
Hose Set Assemblies to Garments
Figure 1 - Flowtron Universal Pump
SER0009
Flowtron Universal Service Manual
Run/Standby Button
Issue 2
Sep 2009