Operation; Start Up; Shut Down - Arjohuntleigh Flowtron Excel Instructions For Use Manual

Hide thumbs Also See for Flowtron Excel:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

5. Operation

Start Up

Shut Down Turn the power switch to the off (
Connect the pump to the mains power supply using the
power cable provided. Turn the mains power switch to
the on (
) position and it will illuminate green. The pump
I
performs a two-second self test cycle where the pressure
display, LEDs and alarm are tested. The pump then
proceeds directly to the inflation cycle.
The garments will inflate alternately. The first garment
inflates for approximately 12 seconds and is deflated for
approximately 48 seconds.
The second garment, if used, inflates 30 seconds after
the first garment has deflated and follows the same
inflation/deflation cycle.
If a single garment is attached, press the
button
to prevent the alarm system from indicating
a fault. The system responds with a 'beep' and the
yellow LED on the Front Panel illuminates.
If the
Single Leg
garments are connected, the system will
automatically reset to two-garment operation
after two single-garment inflation cycles.
Verify that the pressure display is indicating the desired
output pressure prescribed by the physician. Refer to
section "Pressure Adjustment" on page 8 for specific
pressure setting instructions.
If the operation or performance of the pump
changes during use, refer to "Troubleshooting -
Alarm Condition" on page 11 of this IFU before
calling a service engineer or contacting your
local ArjoHuntleigh sales office.
Loss of mains power will halt therapy.
the power off will stop the patient therapy.
If it is required to completely isolate the pump
from the mains power, remove the plug from the
mains power socket.
Disconnect and remove the garment(s) as required.
Garments are for single patient use only. Do not
use the garments on a different patient after
treatment.