Advertisement

Century
INSTRUCTIONS FOR USE
0086
...with people in mind
04.AK.02_10GB. January 2015

Advertisement

Table of Contents
loading

Summary of Contents for Arjohuntleigh Century

  • Page 1 Century INSTRUCTIONS FOR USE 0086 ...with people in mind 04.AK.02_10GB. January 2015...
  • Page 2 Mandatory to read the Instructions for Use Design Policy and Copyright ® and ™ are trademarks belonging to the ArjoHuntleigh group of companies. © ArjoHuntleigh AB 2014. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice.The content of this publication may not be copied either whole or in...
  • Page 3: Table Of Contents

    Cleaning and Disinfection Instructions ........21 Care and Preventive Maintenance .......... 29 Troubleshooting ..............33 Technical Specifications ............37 Labels ..................40 Approvals and List of Standards ..........42 Electromagnetic Compatibility ..........43 Parts and Accessories ............46 ArjoHuntleigh Addresses ..........Last Page...
  • Page 4: Foreword

    Failure to follow these instructions may cause damage Please read this Instructions for Use (IFU) in its entirety to all or parts of the system or equipment. before using your Century bath. Information in this IFU is crucial to the proper operation and maintenance of the NOTE equipment, will help protect your product, and ensure that the equipment performs to your satisfaction.
  • Page 5: Intended Use

    Expected Lifetime This bath is intended for assisted bathing and showering of adult residents in hospital The expected life of the Century is unless otherwise and care facilities. Century must be used by stated ten (10) years, if subject to preventive...
  • Page 6: Safety Instructions

    Safety Instructions WARNING WARNING To avoid falling, make sure that the patient To avoid hair entrapments under water, is positioned correctly and that the safety always install the Hydromassage system belt is being used, properly fastened and suction cover before use. tightened.
  • Page 7: Preparations

    Actions Before Every Use Site Preparations (5 steps) Prior to the installation of the Century bath, the floor and walls within the required area must be finished. 1 Check that all parts are in place. Compare to section Parts Designation on page 8.
  • Page 8: Parts Designation

    Parts Designation 1 Bathtub 2 Leg 3 Foot 4 Control panel 5 Outlet for bathtub filling 6 Patient shower handle 7 Popup-drain 8 Sensors 9 Surface overflow outlet and drain manoeuvring knob 10 Dispenser for Bath oil or Arjo Sound Water Conditioner (for Hydrosound system) (optional) 11 Shampoo dispenser (optional) 12 Disinfectant spray handle (behind door)
  • Page 9 left right...
  • Page 10: Control Panel

    Control Panel Temperature display Patient shower Bathtub water Shampoo & Body wash Temperature setting Temperature display Fill and shower Bath Oil Hydromassage/ Tub fill Disinfection / Autoclean or ArjoSound Hydrosound System Auto fill Water Conditioner Buttons and Displays Shampoo & Body Wash Button Press the Shampoo &...
  • Page 11 Patient Shower Button (2 steps) 1 Press the Patient Shower button to start the shower. 2 Press the Patient Shower button again to stop the shower. The LED lights up when active. The shower function automatically shuts off after 15 minutes. Press the Patient Shower button to restart.
  • Page 12: Product Description/Functions

    The bath system is equipped with an auto fill function, which fills the bathtub to a preset water level. For adjustments or assistance, call your ArjoHuntleigh Service Department. 1 To start to fill the bathtub, press the Tub Fill/Auto Fill button.
  • Page 13: Scalding Protection

    Scalding Protection Century bath is equipped with an electric scalding protection to reach a maximum safety for the patient. The function is used during filling, showering and cleaning. The scalding protection is activated when the water temperature gets too high. If the bathtub water reaches 45 °C (113 °F) or...
  • Page 14: Disinfection Unit

    WARNING To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient. Do note use gloves as it may insulate to the extent that the water temperature can be misjudged. Point the flow of the water away from the patient.
  • Page 15: Hydromassage™ System

    Hydromassage™ System Fig. 1 WARNING To avoid hair entrapments under water, always install the Hydromassage system suction cover before use. Use the Hydromassage™ System (5 Steps) 1 Before the hydromassage starts, fill the bathtub over the Hydromassage system outlet (A), according to section Bathe Patient on page 19.
  • Page 16 Hydromassage with Treatment Hose (7 optional steps) NOTE Fig. 1 If hydromassage is active, stop it by pressing the Hydromassage system button on the control panel before connecting the treatment hose. 1 Remove the Hydromassage system nozzle by pulling it out. 2 Connect the treatment hose to the quick coupling in the bathtub.
  • Page 17: Hydrosound® System

    Hydrosound® System ArjoHuntleigh Hydrosound® system is a method of cleaning patients by means of low frequency, low intensity ultra sound. Working together with water, it is an effective and skin friendly method of cleaning the skin. The amount of time and intensity setting used for Hydrosound system cleaning is a matter of choice consistent with the degree of patient cleanliness required and patient comfort.
  • Page 18 Fig. 3 5 It is possible to select different intensity levels by turning the knob on the upper right side of the bathtub. (See Fig. 3) • The intensity is increased by turning the intensity knob clockwise. • The levels 20, 40, 60, 80 or 100 (mW/cm ) and Test can be chosen step by step.
  • Page 19: Bathe Patient

    Bathe Patient Transfer and Bathe the Patient Fig. 1 (14 steps) 1 Close the drainage plug by turning the drain control knop counter clockwise. 2 Press the Tub Fill button to fill the bathtub WARNING Fig. 2 To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient.
  • Page 20 Fig. 5 6 Position the lift hygiene chair next to the end of the bath. WARNING To avoid entrapment or pinching of genitals, make sure there is enough clearance during movement over the tub edge, toilet, bedpan or other furniture. Raise the lift hygiene chair to a position where it is clear from the edge of the bath and move the patient over.
  • Page 21: Cleaning And Disinfection Instructions

    For daily disinfection we recommend Arjo Clean WARNING (Arjo Cen-Kleen IV), especially developed for use with ArjoHuntleigh Bath Systems, in order to To avoid eye or skin irritation, never avoid scratches in the plastic surface. disinfect in the presence of a patient.
  • Page 22 • Before every use, make sure the product is clean and disinfected. Do not forget to contact ArjoHuntleigh for the annual Preventive Maintenance visit.
  • Page 23 2 Open the access door on the left side of the panel. (See Fig. 1) 3 Disconnect the suction part (A) from the disinfectant bottle. 4 Discard the old bottle and replace with a new ArjoHuntleigh Disinfectant bottle. 5 Connect the suction part (A) to the new bottle.
  • Page 24 Clean and Disinfect with Disinfection Spray Handle or Spray Bottle (Still Bath and Hydrosound® System) (42 steps) Fig. 1 Before every use, make sure that the bath system is clean and disinfected. This procedure cleans and disinfects the bath system. Remove visual residue (Step 1-6 of 42) 1 Check the level of disinfectant in the disinfectant bottle.
  • Page 25 Repeat rinse if needed. 26 Examine bathtub and accessories for visible damage. For Century bath with Hydromassage system, go to section Clean and Disinfect Hydromassage™ System with Automatic Disinfection on page 27, after that proceed with Disinfect (Step 27-42 of 42) on page 26.
  • Page 26 Disinfect (Step 27-42 of 42) 27 Open the access door on the left side of the panel and lift out the disinfection spray handle or use a disinfection spray bottle. 28 For disinfection spray handle only: Direct the disinfectant spray towards the bottom of the bathtub and press the Disinfection button on the control panel.
  • Page 27 40 Wet a new cloth with water and remove all traces of disinfectant on contact areas. The cloth should occasionally be rinsed in running water when removing disinfectant. 41 Leave to air dry. 42 Attach any accessories. NOTE Let the pillow dry with the suction cups upwards. Fig.
  • Page 28 Fig. 4 8 Open the access door on the left side of the panel and lift out the disinfection spray handle or use a disinfection spray bottle. 9 For disinfection spray handle only: Direct the disinfectant spray towards the bottom of the bathtub and press the Disinfection button on the control panel.
  • Page 29: Care And Preventive Maintenance

    Care and Preventive Maintenance The Century bath is subject to wear and tear, and the following maintenance instructions must be performed, when specified, to make sure the equipment remains within its original manufacturing specification. WARNING To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow the recommended maintenance schedule.
  • Page 30 Between Patients • Clean and disinfect: Before every use, make sure the Century bath is cleaned and disinfected according to section Cleaning and Disinfection Instructions on page 21. Every Day Check liquid levels: Check liquid levels in the bottles of disin- fectant, shampoo and bath oil (ArjoSound water conditioner).
  • Page 31 Every Month Fig. 1 • Check and clean the shower head (5 steps): Remove and clean the strainers in the shower head. 1 Loosen the screw. (See Fig. 1) 2 Remove all loose parts (A) according to the illustration. 3 Clean all parts carefully. 4 Reassemble the shower head.
  • Page 32 Yearly Checks by Qualified Personnel Only The Century bath must be serviced once a year in accordance with the Maintenance and Repair Manual. To make sure that you have problem-free and safe operation, it is recommended to have regular scheduled services with calibration of the thermostatic mixers by a trained technician.
  • Page 33: Troubleshooting

    Troubleshooting NOTE Before starting the troubleshooting check: Power supply, Water supply and the LEDs on control panel. If the problem can not be solved with the problem solving actions below, please contact an ArjoHuntleigh authorized technician. General Problems PROBLEM ACTION The water is too hot or cold when filling the bath or •...
  • Page 34 PROBLEM ACTION Water temperature fluctuates. • Probably caused by supply temperature fluctuation. • In order to feel if any hot water is coming in, feel the incoming hot water line to see if hot. • Check and clean the incoming water filters. See Care and Preventive Maintenance on page 29.
  • Page 35 Control Panel Problems PROBLEM ACTION No shampoo, body wash, bath oil/ArjoSound water • Check power supply and fuses. conditioner emerges from the dispenser when pressing • Check level in Shampoo/Bath oil container. the Shampoo & Body Wash or Bath Oil/ArjoSound •...
  • Page 36 PROBLEM ACTION Hydromassage/Hydrosound function does not start. • Check power supply and fuses. • Make sure that no other function is activated. • Make sure that the water level is above the top level sensor in the bathtub. • Check the Hydrosound system LED, a yellow flash indicates that a problem has occurred.
  • Page 37: Technical Specifications

    Technical Specifications Technical Data: Still Bath, Hydrosound® System & Hydromassage™ System NOTE All drainage, water and electrical installation work should be carried out by qualified personnel, in accordance with local and national regulations. Power Market Voltage Frequency Fuse (slow blow) consumption Europe 230/240 V single...
  • Page 38 Sound Levels (LwA, maximum) Sound level at hydromassage 76 dBA Sound level at bathtub filling 80 dBA Water Supply Cold water G3/4” male (customers connection) Warm water G3/4” male (customers connection) Connect with shut off valves between unit and supply net Minimum through Ø...
  • Page 39 This requirement applies in the European Union. Do not dispose these products as unsorted municipal waste. You can return your device and accessories to ArjoHuntleigh or you can contact your local authorities for advice on disposal.
  • Page 40: Labels

    Labels Label Explanation Manufacturing date. Identification States the product identification; Label serial number, year and month of manufacturing. 4ML5 Certificate control number for UL Label for equipotential connection For example, 1402XXXXX where point (bonding lug). “14” states manufacturing year, “02” states manufacturing month and “XXXXX”...
  • Page 41 Attention Label placed on the disinfectant spray handle Entrapment Hydrosound System warning warning Label Applied part IFU Label Data Label UL classification Label Field wiring box Identification Label Label Water connection Equipotential Label Label...
  • Page 42: Approvals And List Of Standards

    Approvals and List of Standards Plumbing Approval Standard Description ANSI, Z124.1.2–2005 Plastic Bathtub and Shower Units ANSI/ASME, A112.19.8–1987 Suction Fittings ANSI/ASME, 1001–2008 Atmospheric Type Vacuum Breakers ANSI/ASME, 1014–1989 Backflow Prevention Devices for Hand-Held Showers ANSI/ASME, 1016–2005 Automatic Compensating Valves for Individual Showers and Bathtub/Shower Combinations ANSI/ASME, 1056–2001 Spill Resistant Vacuum Breakers...
  • Page 43: Electromagnetic Compatibility

    EMC information below. The Century bath is intended for the use in the electromagnetic environment specified below. The customer or the user of the Century bath should assure that it is used in such an environment.
  • Page 44 Power frequency 3 A/m 0.3 A/m If functional disturbance occurs on the (50/60 Hz) Century bath it may be necessary to install magnetic shielding. The power Magnetic field frequency magnetic field should be IEC 61000-4-8 measured prior to installation to assure that it is sufficiently low.
  • Page 45 RF communications equipment (transmitters) and the Century bath by maintaining a distance of not less than 1 meter to any part of the Century bath, including cables, if the transmitters output power rating exceeds 1 W.
  • Page 46: Parts And Accessories

    Parts and Accessories 8341958 ARA0500-03 ARA0007-01 Dual Hydromassage Hydromassage Tray Nozzle Treatment Hose (for bath utensils) For Disinfectant contact SRB1000 SRB5000 your local representative ArjoSound™ Arjo Care™ Water Conditioner (Shampoo, Body Wash) (for Hydrosound System) 4 X 2 Litre bottles 2 X 2 Litre bottles SRB4000 Arjo Oil™...
  • Page 47: Arjohuntleigh Addresses

    Free: +1 800 072 040 Fax: +33 (0) 3 20 28 13 14 POLSKA Fax: + 61 89337 9077 E-mail: info.france@arjohuntleigh.com ArjoHuntleigh Polska Sp. z o.o. BELGIQUE / BELGIË HONG KONG ul. Ks Piotra Wawrzyniaka 2 ArjoHuntleigh NV/SA ArjoHuntleigh (Hong Kong) Ltd...
  • Page 48 GETINGE GROUP is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh focuses on patient mobility and wound management solutions. GETINGE provides solutions for infection control within healthcare and contamination prevention within life sciences.

Table of Contents