Safeline Signature Operating And Instruction Manual page 98

Table of Contents

Advertisement

3)
Press the soft key adjacent to tm1
4)
Press the soft key adjacent to REJECT TIME
5) Use the cursor Up and Down keys to adjust the
Reject Time to the required value.
6) Press the Enter key to accept the new settings.
7) Press the Enter key a second time to proceed to
the next display.
8)
The display shown opposite will appear for
five seconds before returning to the Engineer
Mode
display
Testing the Reject Device.
9) Check the reject device operates correctly with different
sizes of metal contaminant.
10) If for any reason the setting of the tm1 timer is
not correct return to step 1) and repeat all the
steps.
82
OPERATING INSTRUCTIONS
4618 - 021