Safeline Signature Operating And Instruction Manual page 147

Table of Contents

Advertisement

Command Identifier - 'j'
procedure
The Automatic sensitivity limit procedure can be started remotely by the host or from the front panel keypad, in either
case the front panel display and the communications commands would be in synchronization. When the host starts
the Automatic sensitivity limit procedure remotely, with the command 'j1'given below, then the display will operate
just as if the Automatic sensitivity limit procedure had been started from the front panel keyboard.
Commands the host may transmit:
The host can send any of the following messages :-
1,2,3
',j1'
This starts the automatic sensitivity limit procedure operating for the current product number.
'j?'
This returns the current state of the automatic sensitivity limit procedure (see the 'j'n command
description below).
'j99'
This command is sent when the host wishes to abort the automatic sensitivity limit procedure and return
to normal operation. This is equivalent to pressing the front panel Security key during the automatic
sensitivity limit, and will cause the same action to occur in the metal detector.
Commands the metal detector may transmit
The following command will be transmitted in response to a 'j?' command above, or automatically during the
operation of automatic sensitivity limit.
'j' n
n is a number indicating the state of the automatic sensitivity limit's progress, the number, n,
corresponds with the displays on the front panel display as shown below.
Note the special case n = 0 means that automatic sensitivity limit is not running.
Changing the frequency
passing through aperture
Notes
1
During an automatic sensitivity limit procedure, initiated by the host, the transmission of the bargraph and
detection messages ('A'n and 'R'n), will be suppressed. After completing the automatic sensitivity limit, these
messages will be resumed if they had been enabled prior to entering the automatic sensitivity limit.
2
The command 'j1' cannot be sent if the current product number is 0. Automatic sensitivity limit is not available
on product number 0.
3
The command 'j1' cannot be sent if either Automatic Speed or Automatic setup is running.
4618 - 021
please wait ...
Ensure no product
please wait ...
Automatic sensitivity limit
'j1'
'j2'
'j5'
APPENDIX D
131