OPERATING INSTRUCTIONS
If product signal cancellation is required for the
current product number select YES, else select NO.
This display appears when automatic setup has
finished. The display will be seen for a few seconds,
and then will revert back to the display from which
automatic setup was started.
Passing the product through the aperture should now
give 2 to 4 segments on the bar graph display.
NOTE
In a minority of applications it may be necessary to
manually trim the phase and/or sensitivity to optimise
performance. If this is required refer to the 'How to
Manually Trim The Phase And/Or Sensitivity' later in
this section.
Test the sensitivity of the detector by passing metal
samples through the aperture of the detector
(preferably on or in the product).
If required manually adjust Sensitivity (page 53) to
achieve the required spherical performance standard,
i.e. 1.0, 1.5, 2.0 Fe, etc.
Note the sensitivity level for the bargraph may be
limited by the vibration characteristics of the system.
If this is the case then ensure that manually
increasing the sensitivity does not cause the detector
to trigger from system vibration.
This display occurs if automatic setup is being
performed on a product number with the inverse
detect feature enabled.
Selecting YES allows the calibration of the product
and additive combination to be performed again.
This allows different orientations of the additive to be
tested - the automatic setup will always choose the
best setting out of all of the orientations tried.
When there are no more orientations to be tried select
NO.
4618-021
51