Parameter Assignment; Overview Of Parameters And Functions - Siemens SITRANS P Operating Instructions Manual

Pressure transmitter
Hide thumbs Also See for SITRANS P:
Table of Contents

Advertisement

Parameter assignment

9.1

Overview of parameters and functions

Introduction
You can operate the device via local operation or remote operation (e.g. HART communicator,
SIMATIC PDM).
● The parameters that you can reach over the device with a display are marked by the
parameter ID. Hereinafter, the parameter ID is always written inside parentheses after the
parameter name. Example: Parameter "Damping value" [04].
● You can access the complete number of parameters via remote operation.
The device-specific parameters are available in each tool for configuration.
The instructions or online help for these tools will provide you with information on how to use
the different tools for parameter assignment.
List of parameters and functions
The following parameters are available via the local operation and via remote operation (e.g.
SIMATIC PDM).
The parameters are grouped according to their function in the following overview:
Quick start
Quick start wizard
Current output
Apply lower range value (with pressure applied)
Apply upper range value (with pressure applied)
Set lower range value (without pressure applied)
Set upper range value (without pressure applied)
Set damping value
SITRANS P320/P420 with 4 to 20 mA/HART
Operating Instructions, 11/2019, A5E44852162-AC
SIMATIC PDM
Device with display (lo‐
cal operation)
Menu command "De‐
-
vice > Wizard - Quick
start..."
SIMATIC PDM
Device with display (lo‐
cal operation)
Menu command "De‐
Apply lower range val‐
vice > Apply values"
ue parameter [08]
(Page 122)
Apply upper range val‐
ue parameter [09]
(Page 122)
"Settings > Current out‐
Set lower range value
put" parameter group
parameter [02]
(Page 108)
Set upper range value
parameter [03]
(Page 108)
"Settings > Current out‐
Damping value [04]
put" parameter group
(Page 110)
9
Device without display
(local operation)
-
Device without display
(local operation)
Commissioning the de‐
vice without display
(Page 11)
-
-
101

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans p320Sitrans p420

Table of Contents