Installing electrical connections and cable entries

The device is delivered with dust caps installed on both sides at the factory.
You use the order options A.. to define the type of electrical connections and cable entries
(cable gland, sealing plug or device plug) for your device.
These components are delivered with the device.
● To order the device with installed electrical connections and cable entries, select an
additional order option A.. for the installation (e.g. device plug mounted on the right).
Procedure
For the first installation follow these steps:
1. Ensure that the seals are clean and undamaged.
2. To ensure the IP degree of protection and explosion protection of the pressure transmitter,
close the cable entries with a sealing plug, a cable gland or a device plug.
See also
Replacing electrical connections and cable entries (Page 185)
5.6

Rotating the display

To read the display in any mounting position, you have the option of gradually rotating the
display 360°.
Procedure
1. De-energize the device.
2. If available, loosen the front safety catch with a 3 mm Allen key.
3. Unscrew the front cover.
SITRANS P320/P420 with 4 to 20 mA/HART
Operating Instructions, 11/2019, A5E44852162-AC
Start of scale value
Full-scale value
Density of the process medium in the container
Density of the filling oil in the capillary line of the remote seal
Acceleration due to gravity
Lower filling level
Upper filling level
Gland distance
Distance container flange to pressure transmitter
2
Installing/mounting
5.6 Rotating the display
61