Limits - ILX Lightwave LDT-5900 Series User Manual

High power temperature controller, precision temperature controller
Table of Contents

Advertisement

The coefficients can be modified by using the Adjust Knob and then pressing the
ENTER / LOCK key. For more details on sensor, see "Sensor Options" later in this
chapter.

LIMITS

The Limits menu allows the user to select and modify the setpoint and error limits.
The arrow keys and adjust knob can be used to cycle through and select or
modify the following parameters:
When the limit parameter is selected, the display will look as follows.
When these limits are exceeded, the corresponding LIMIT or ERROR LED in the
Display Section is illuminated. Als0, exceeding these limits may cause the
controller output to be turned off automatically. (See the Output Off Registers
section in Chapter 3).
01_10
LDT-59XX
C1: [001.125]
C2: [002.347
Temperature High / Low Limits
ITE (TE Current) High / Low Limits
Sensor High / Low Limits
LDT-59XX
Temp High:[060.000]
Temp Low: 000.000
TEMPERATURE CONTROLLER
VTE (TE Voltage) High / Low Limits
Set Point Tolerance (for Trigger Out)
TEMPERATURE CONTROLLER
O P E R A T I O N S
Front Pa nel Opera t ion
LDT-5980/ 5948
2
C H A P T E R
27