C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
7j
Enable Cal Mode by entering the remote command MODE CAL
7k
Enter the new calibration values using the command CAL:RAC m
NOTE: The offset value that must be entered in the above step is the negative of the
value calculated in step i.
7l
Save the new values to flash memory by using the CAL:SAVE command.
7m
Return to R
mode by the MODE RAC command.
AC
7n
Verify the calibration values are correct by issuing the query CAL:RAC? to check
them.
Ca libra t ion Procedure
, -b
RAC
RAC
LDT-5980/ 5948
5
C H A P T E R