Snsr; Ite Or Vte - ILX Lightwave LDT-5900 Series User Manual

High power temperature controller, precision temperature controller
Table of Contents

Advertisement

2
C H A P T E R
T
Selecting T will enable the instrument to operate in constant temperature mode
when OUTPUT is turned on. In T mode, the controller output current to the TEC is
continuously adjusted to maintain the setpoint temperature at the sensor. Any of
the sensor types (thermistor, RTD, IC-I, IC-V) can be used in this control mode.
Auto-Tune and user selectable PID constants are active in this mode.
R

SNSR

Selecting R
when OUTPUT is turned on. In R
TEC is continuously adjusted to maintain the setpoint sensor resistance
(thermistor, RTD), current (IC-I) or voltage (IC-V). Auto-Tune and user selectable
PID constants are active in this mode.
I
TE
Selecting I
or constant TEC voltage mode when OUTPUT is turned on. In I
setpoint current is continuously output to the TEC.
30
LDT-5980/ 5948
O P E R A T I O N S
Front Pa nel Opera t ion
LDT-59XX
0[035.000]
will enable the instrument to operate in constant sensor mode
SNSR
LDT-59XX
[009.988]
0
or V
TE
or V
will enable the instrument to operate in constant TEC current
TE
TE
TEMPERATURE CONTROLLER
o
C
mode, the controller output current to the
SNSR
TEMPERATURE CONTROLLER
k
mode, the
TE

Advertisement