C H A P T E R
ENABle:OUTOFF
Action
Parameters
Output Off Enable Register 1
Bit Number
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
* This bit cannot be cleared.
82
LDT-5980/ 5948
C O M M A N D R E F E R E N C E
Com ma nd Reference
The ENABle:OUTOFF command sets the controller Output Off Enable Registers. The
bits in these registers specify the conditions that will cause the controller to automatically
turn off.
Two <nrf values> that represent the sum of the enabled bits for each register. The first
represents the sum of the enabled bits for register 1 and the second represents the sum
of the enabled bits for register 0.
(e.g. ENAB:OUTOFF <Reg 1>, <Reg 0>)
1 = Enabled
Condition
Reserved
Reserved
Reserved
Reserved
Measurement Out of Setpoint Tolerance
Reserved
Reserved
Reserved
Reserved
Internal PCB Temperature Out of Tolerence*
Reserved
Reserved
TEC Current Runaway
Reserved
Reserved
Reserved
0 = Disabled
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
Default Value
Decimal Value
0
1
0
2
0
4
0
8
0
16
0
32
0
64
0
128
0
256
1
512
0
1024
0
2048
0
4096
0
8192
0
16384
0
32768