C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
5
C H A P T E R
Ca libra t ion
location as the remaining output wires. Failure to do this will result in voltage
differentials that will degrade the calibration accuracy. Finally, pins 4 and 17 and 5
and 18 must be shorted together to allow operation at maximum current.
A computer capable of GPIB or RS232 communications is required in order to
calibrate the instrument. CALIBRATION VIA THE FRONT PANEL IS NOT
POSSIBLE. It shall be assumed that the computer and instrument are already
connected and communicating correctly.
119