ILX Lightwave LDT-5900 Series User Manual page 88

High power temperature controller, precision temperature controller
Table of Contents

Advertisement

4
C H A P T E R
CAL:SENsor:VOLTage:1MA?
Action
Response
Notes
Example
CAL:SOURCE:SENsor:10UA
Action
Parameters
Notes
Example
CAL:SOURCE:SENsor:10UA?
Action
Response
Notes
Example
72
LDT-5980/ 5948
C O M M A N D R E F E R E N C E
Com ma nd Reference
The CAL:SENsor:VOLTage:1MA? query returns the calibration coefficients for the sensor
voltage measurement with a 1 mA sensor supply current.
The response is two values; the first represents the calibration voltage measurement
slope and the second represents the voltage offset coefficients.
See the Calibration section of Chapter 5 for complete calibration instructions.
CAL:SEN:VOLT:1MA? Response: 1,0 means a slope of 1.0 V/V and an offset of 0 V for
the sensor voltage measurement calibration.
The CAL:SOURCE:SENsor:10UA is used to enter the calibration coefficient for the 10
µA sensor current source.
One <nrf value> is required representing the 10 µA scale factor.
This is a service related command. Calibration mode should not be entered unless the
user has the proper measurement equipment in place. See the Calibration section of
Chapter 5 for complete calibration instructions
CAL:SOURCE:SEN:10UA 0.9217 Action: Scales the original sensor output current at the
10 µA setting by a factor of 0.9217.
The CAL:SOURCE:SENsor:10UA? query returns the calibration coefficient for the 10 µA
sensor current source.
The response is one value that represents 10 µA current source calibration coefficient.
See the Calibration section of Chapter 5 for complete calibration instructions
CAL:SOURCE:SEN:10UA? Response: 0.93, means a scalefactor of 0.93 has been
applied for the 10 A current source calibration.
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ldt-5980Ldt-5948

Table of Contents