LIMit:SENsor:HIgh
Action
Parameters
Notes
Example
LIMit:SENsor:HIgh?
Action
Response
Notes
Example
92
LDT-5980/ 5948
C O M M A N D R E F E R E N C E
Com ma nd Reference
The LIMit:SENsor:HIgh command sets sensor upper limit value.
An <nrf value> that represents the upper sensor limit. Units depend on which sensor is
selected. For thermistor and RTD units are Ohms, for ICI Amps, and for ICV Volts.
Acceptable values for thermistors are 1 to 600000 .
Acceptable values for RTD are 0.1 to 60000.0 .
Acceptable values for ICI are 0.000010 to 0.000600 Amps.
Acceptable values for ICV are 0.100 to 6.000 Volts.
The minimum upper limit value allowed will be determined by the value set with the
LIMit:SENsor:LOw command
By default, the controller output is turned off when the measured sensor exceeds this
setting in Rsnsr mode. This feature can be disabled by clearing bit 11 in the Output Off
Enable Register 0, using the ENABle:OUTOFF command.
LIM:SEN:HI 450.100 action: Sets the sensor upper limit to a maximum of 450.1 .
The LIMit:SENsor:HIgh? query returns the value of the sensor upper limit setting.
The response is a single value that represents the upper limit and the units depend on
which sensor is selected. For thermistor or RTD the value will be in Ohms, for ICI Amps,
and for ICV Volts.
This limit is only in effect in R
LIM:SEN:HI? response: 450100.00000000 when the thermistor is selected means a
upper sensor limit of 450,100 Ohms.
mode.