C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Ca libra t ion Procedure
6af
Calculate the 1 mA Sensor Voltage Measurement slope and offset using the
equations below:
m
vsens
b
vsens
6ag Enable Cal Mode by entering the remote command MODE CAL.
6ah Enter the new calibration values by using the command
CAL:SENsor:VOLTage:1MA m
6ai
Save the new values to flash memory by using the CAL:SAVE command.
6aj
Return to Sensor Mode by using the MODE SENSOR command.
6ak Verify the correct storage of the calibration values by entering
CAL:SENsor:VOLTage:100UA?
RAC Calibration
7a
NOTE: R
mode and the subsequent recalibration thereof is only possible with
AC
firmware version 01.03.09 and above.
7b
Select R
mode and connect TE output across TE Cal resistor R
AC
highest possible accuracy, ensure TE voltage sense lines are connected to the same
points on the TE Cal resistor as the TE output.
7c
Enable R
measurement by sending the remote command OUTput ON
AC
7d
Record the first measurement for R
MEASure:RAC?
7e
Re-enable R
measurement with another OUTput ON command followed by a
AC
MEASure:RAC? query and record this second R
7f
Repeat this process several additional times to generate a well-defined mean value
for R
.
AC1
7g
Replace R
with R
TE
7h
Take several measurements of R
generate a well-defined mean value for R
7i
Calculate the R
Slope and Offset values using the following formulae:
AC
m
b
RAC
V
V
sens
3
b
sens
3
a
3
V
V
3
b
3
a
V
m
V
3
sens
3
b
vsens
3
3
b
, b
vsens3
vsens3
by querying the instrument with
AC1
Cal resistor R
.
AC
RAC
using OUTput ON and MEASure:RAC? to
RAC
.
AC2
R
R
RAC
TE
RAC
R
R
AC
2
AC
1
R
m
R
RAC
RAC
AC
2
.
. To ensure
TE
measurement.
AC1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ldt-5980Ldt-5948

Table of Contents