Table 5.1 Problem and Action
SYMPTOM

Error Codes

The LDT-5900 indicates general TEC operational error conditions on the display
or through the GPIB using the ERR? command. This section contains descriptions
of the errors that are specific to the LDT-5900.
Front panel Error Indicators
When an error occurs, the error code and message appears on the front panel
display.
Table 5.2 LDT-5980/5948 Error Codes
Range
E-100 to E-199
E-200 to E-299
E-300 to E-399
E-400 to E-499
E-700 to E-799
E-800 to E-899
Error Code
0
123
124
125
01_10
C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Output turns off or turns off intermittently.
Cannot increase current limit to 5A/10A.
Area of Operation
Command / Parser Errors
Execution Control Errors
Remote Operation Errors
TEC Control Errors
System Errors
Internal Errors
Explanation
No errors found
Command not found
Look-up failed because query / command match failed
Parser syntax error
Troubleshoot ing
CORRECTIVE ACTIONS
Check that the AC power cord connection is secure.
Power-line dropouts may reset the unit and when
power is restored, the output will be off.
Check the TE module connections. A high
impedance on the TE load can cause the output to
exceed the compliance voltage momentarily. If
enabled with GPIB, this condition will shut off the
output. There are other events or conditions that will
turn the output off if enabled with GPIB. See Output
Off Registers in Chapter 3 for more information.
Check the sensor connections at the module and the
load.
Changing control modes while the output is on will
turn the output off.
Verify output cable is securely connected to output
connector and that the Cable ID jumpers are correct.