ILX Lightwave LDT-5900 Series User Manual page 87

High power temperature controller, precision temperature controller
Table of Contents

Advertisement

CAL:SENsor:VOLTage:100UA
Action
Parameters
Notes
Example
CAL:SENsor:VOLTage:100UA?
Action
Response
Notes
Example
CAL:SENsor:VOLTage:1MA
Action
Parameters
Notes
Example
01_10
The CAL:SENsor:VOLTage:100UA command is used to enter the calibration coefficients
for the sensor voltage measurement with a 100 A sensor supply current.
Two values are required to set the sensor voltage calibration. The first represents the
slope and the second reprsents the offset in voltage measurement.
This is a service command. Calibration mode should not be entered unless the user has
the proper measurement equipment in place. See the Calibration section of Chapter 5 for
complete calibration instructions.
CAL:SEN:VOLT:100UA 1,0 action: Enters a slope of 1.0 V/V and an offset of 0 V for the
sensor voltage measurement calibration.
The CAL:SENsor:VOLTage:100UA? query returns the calibration coefficients for the
sensor voltage measurement with a 100 A sensor supply current.
The response is two values; the first represents the calibration voltage measurement
slope and the second represents the voltage offset coefficients.
See the Calibration section of Chapter 5 for complete calibration instructions.
CAL:SEN:VOLT:100UA? Response: 1,0 means a slope of 1.0 V/V and an offset of 0 V for
the sensor voltage measurement calibration.
The CAL:SENsor:VOLTage:1MA command is used to enter the calibration coefficients
for the sensor voltage measurement with a 1 mA sensor supply current.
Two values are required to set the sensor voltage calibration. The first represents the
slope and the second represents the offset in voltage measurement.
This is a service command. Calibration mode should not be entered unless the user has
the proper measurement equipment in place. See the Calibration section of Chapter 5 for
complete calibration instructions.
CAL:SEN:VOLT:1MA 1,0 action: Enters a slope of 1.0 V/V and an offset of 0 V for the
sensor voltage measurement calibration
C O M M A N D R E F E R E N C E
Comm a nd Reference
4
C H A P T E R
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
71
LDT-5980/ 5948

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ldt-5980Ldt-5948

Table of Contents