SENsor
Action
Parameter
Notes
Example
SENsor?
Action
Response
Notes
Example
SET:
106
LDT-5980/ 5948
C O M M A N D R E F E R E N C E
Com ma nd Reference
The SENsor commands selects the sensor type and sense current.
One <character data value> representing the sensor type is required. The values
accepted are THERM10UA, THERM100UA, THERM1MA, ICI, ICV, RTD10UA,
RTD100UA, or RTD1MA.
For thermistor or RTD sensors, a sense current is applied through the sensor, and the
resulting voltage is measured by the instrument. The measured voltage range is 0-6V.
The best sense current to use depends on the expected resistance at the setpoint. The
largest possible voltage swing in the vicinity of the setpoint will result in the best stability.
For more details regarding proper sensor selection and sense current, refer to Chapter 2.
SEN THERM100UA action: sets the sensor to thermistor and the sense current to
100 µA.
The SENsor? query returns the selected sensor type and sense current (if appropriate).
The response is a single character string that represents the selected sensor and is one
of the following:
THERM10UA, THERM100UA, THERM1MA, ICI, ICV, RTD10UA, RTD100UA, or
RTD1MA.
SENsor? Response: ICI, means the ICI sensor is selected.
The SET: command path is used to access the control mode setpoint commands and
queries.
The following may be reached directly from the SET: command path:
SET:ITE
SET:SENsor
SET:TEMP
SET:VTE
SET:ITE?
SET:SENsor?
SET:TEMP?
SET:VTE?
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL
C
OMMON
D
D
EVICE
EPENDENT
F
P
RONT
ANEL

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ldt-5980Ldt-5948

Table of Contents